Bench pH, ORP and Temperature Meter with GLP
Benchtop pH, ORP and temperature meter with GLP functions, USB and printer output for accurate, traceable laboratory results.
The AK131 Max is the ideal choice for laboratories that require a robust, high-performance instrument with features that ensure accuracy, traceability, and reliable results.
The instrument has GLP (Good Laboratory Practices) functions, which allow you to set the validity of the last adjustment made, check which standards were used, monitor the results obtained, and consult the slope of the pH electrode.


Applications
| Measurement Specifications | |||
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| pH | -2.000 to 19.999 pH | 0.001 / 0.01 pH | ± 0.002 / ± 0.01 pH |
| ORP / Redox Potential | -1999.9 to 1999.9 mV | 0.1 mV | ± 1 mV |
| Temperature | -5.0 to 100.0 °C | 0.1 °C | ± 0.5 °C |
| General Specifications | |
| Electrode operating temperature | 0 to 80 °C |
| Probe operating temperature | 0 to 100 °C |
| PC communication | USB Type A |
| Printer communication | RS232 |
| Display backlight | On/Off |
| Freezing the reading (HOLD) | Manual / Automatic |
| Protection rating | IP65 |
| Operating humidity | 10 to 90 %RH (non-condensing) |
| Power supply | 220 VAC / 12 VDC Adapter |
| Dimensions | 216 x 190 x 59 mm |
| Weight | 950 g (meter only) |
Replacement parts, calibration solutions, and maintenance supplies for the AK131.
Regular cleaning ensures measurement accuracy and extends electrode life. Follow this procedure periodically.
Prepare a small beaker with electrode cleaning solution (pepsin).
Remove the electrode protection and immerse the tip in the solution for 20–25 minutes.
Remove the electrode from the solution and rinse it with distilled or deionized water before and after each procedure.
Always remove excess water with a soft towel or cloth, without rubbing the bulb.
Check for air bubbles in the sensing bulb. If present, gently shake the electrode vertically to remove them.
Store the electrode in AKSO 3M KCl (potassium chloride) solution for 1 hour.
The electrode is now ready for adjustment — always use fresh calibration solutions.
Important
Tell us what you measure and where. Our team will recommend the right instrument and configure your complete kit.
Expert consultation
Speak directly with application engineers
Custom kit configuration
Instrument, probe, solutions, and carrying case
Global shipping
Direct or through authorized distributors