AK90 Tester: pH/Temperature pocket tester
Ideal for quick, accurate checks in the field, the AK90 is also used for analysing water in pools, lithographic processes, boiler and cooling tower, aquaculture, aquarium and nutrients levels in Hydroponics.
Specifications:
- Range: pH: 0.0 to 14.00 pH
- Temperature: 0 to 50 ºC
- Resolution: pH: 0.1 pH
- Temperature: 0.1ºC
- Accuracy: pH: ± 0.1 pH
- Temperature: ± 1ºC
- pH calibration: 1 or 2 points automatic – 3 standard buffers: 4.01, 7.00 and 10.01 pH
- Automatic Temperature compensation: 0 to 50 °C
- pH/Temp electrode: AK1050 (included)
- Temperature probe: Built-in temperature sensor
- Auto shut-off: 10 minutes
- Reading (HOLD): Manual
- Operating Temperature: 0 to 50 °C
- Operating RH: 10 to 90 % RH
- Casing: IP54
- Battery type: 6VDC (4 batteries LR44 included)
- Dimensions (W x L x H): 38 x 188 x 38mm
- Weight: 85g
AK90 TESTER IS SUPPLIED COMPLETE:
- Hard carrying case
- 20 ml calibration solution pH 7.01
- 20 ml calibration solution pH 4.01
- 20 ml calibration solution pH 10.01
- 20 ml KCL electrolyte solution
- 20 mL cleaning solution
- Instruction manual
- 4 x LR44 batteries
Optional accessories:
- AK1050: Replaceable pH/Temp electrode for AK90
- AK4504: pH 4.01 buffer solution 250 mL
- AK4507: pH 7.00 buffer solution 250 mL
- AK4510: pH 10.00 buffer solution 250 mL
- AK4530: KCL Storage solution for pH electrodes - 250 mL
- AK4531: Cleaning solution for pH electrodes - 250 mL
MANTAINING & CLEANING YOUR pH ELECTRODE
To guarantee the quality of the measurements and the durability of the
electrode, clean it periodically:
- Prepare a small beaker with electrode cleaning solution (pepsin).
- Leave the electrode tip immersed in the solution for 30 minutes;
- Rinse the electrode in distilled water;
- Remove excess water using a soft paper towel;
- Store the electrode in KCL-3M potassium chloride solution for 1 hour;
- Your Tester is now ready for calibration;
- Always use fresh calibration solutions;
NOTES:
- NEVER leave the electrode in the cleaning solution for more than
30 minutes.
- NEVER use abrasives (brushes, rough paper, etc.) to clean the electrode.